Flash Point Tester
| PMA 300 | PMA 500 | |
| Temperature range | From above ambient temperature to 370 °C | From above ambient temperature to 410 °C |
| Ignition type | Electric (encapsulated hot wire) | Electric (encapsulated hot wire) Optional gas ignition |
| Cooling after measurement | Standard cooling | Boost mode |
| Barometric pressure correction | Flash point is automatically corrected to barometric pressure | |
| Flash detection | Automatic detection by thermoelement | |
| Sample temperature | Automatic by Pt100 | |
| Safety | Automatic fire-extinguishing system in combination with a unique optical fire detection system Overheat protection, automatic shut-off | Detects a “flash” outside the cup | Automatic fire-extinguishing system in combination with a unique optical fire detection system Overheat protection, automatic shut-off | Detects a “flash” outside the cup User management with different access levels |
| Calibration | Calibration of sample and block temperature, stirrer speed, and internal barometer | |
| Memory | 1 GB space for approx. 50,000 tests and 1,000 users | |
| Interfaces | 2 × USB, 1 × LAN | 4 × USB, 1 × LAN |
| Power supply | AC 100 V - 120 V, 50/60 Hz AC 220 V - 240 V, 50/60 Hz | |
| Heating power | 2 × 330 W | |
| Gas supply | Integrated fire extinguisher: CO2 or N2 inert gas; inlet pressure 400 kPa to 500 kPa | Integrated fire extinguisher: CO2 or N2 inert gas; inlet pressure 400 kPa to 500 kPa Optional gas ignition: 50 mbar of propane or butane |
| Dimensions | 262 mm × 506 mm × 486 mm | |
| Weight | Ca. 15 kg | |